How to choose a lock-in amplifier: five specs that decide the price
A lock-in amplifier is not a case of the more expensive the better. What matters is whether five numbers — frequency range, input noise, dynamic reserve, time constant and channels — match your experiment. This guide uses the real specifications of ten models in our catalog to turn selection into three steps.
Why a lock-in at all
A 1 nV signal fed into a front end with 5 nV/√Hz input noise is smaller than the noise itself, so plain amplification will never reveal it. A lock-in instead moves the signal onto a known frequency and demodulates only a very narrow band around that single point.
The key is equivalent noise bandwidth (ENBW). With a 1 s time constant the ENBW is roughly 0.25 Hz, and noise falls with the square root of the bandwidth ratio — a gain of several hundred times that no low-noise preamplifier can match. Choosing a lock-in therefore means asking how slow your signal is, how large the noise is, and how strong the interference is.
Spec 1 — frequency range: start from your modulation frequency
The starting point is not the widest frequency range on the market, but where your experiment puts the signal. A lock-in performs best in the middle of its range; operating at the extreme ends usually costs you time constant or sensitivity.
- 1 mHz – 102 kHz: OE1022 series, for slow signals such as low-frequency spectroscopy, photoconductivity and thermal conductivity
- 10 μHz – 250 kHz / 1.5 MHz: OE2011 / OE2021 / OE2012 / OE2022, from ultra-low frequency to a few hundred kHz
- DC – 10 MHz / 80 MHz: OE2031 / OE2041, high-frequency modulation and fast scans
- 1 μHz – 400 MHz: OE2052, wideband dual channel with four demodulators
- 1 μHz – 100 kHz / 500 kHz / 1 MHz: OE1311 / OE1351 / OE1301, modular with eight demodulators
Spec 2 — input noise: how small a signal you can resolve
Three tiers in our catalog are worth remembering: 2.2 to 2.5 nV/√Hz (OE2011 / OE2021 / OE2012 / OE2022), 3 nV/√Hz (OE2031 / OE2041 and OE2052), and 5 nV/√Hz (OE1022 series, OE1311 series, OE1201).
The figure only holds at a matched source impedance. With high-impedance sources above 1 MΩ, current noise and the 1/f corner matter more than voltage noise; for low-temperature and low-frequency work, look at 1/f behaviour rather than the headline number quoted near 997 Hz, which may be far from your operating frequency.
Do not read 5 nV/√Hz as "5 nV is the floor". With a 1 s time constant the usable resolution reaches sub-nanovolt levels, because the noise is suppressed by the ENBW rather than set by the front end.
Spec 3 — dynamic reserve: the larger the interference, the more it matters
Dynamic reserve is the ability to demodulate a small signal while a much larger interferer is present. Above 120 dB means the front end does not saturate when the interference is six orders of magnitude larger than the signal; above 130 dB (OE2011 / OE2021 / OE2012 / OE2022) adds another order of magnitude.
Typical pressure points: strong excitation background in laser-induced fluorescence, large drive currents in electrochemistry, heater PWM interference inside a cryostat. In all three, losing 10 dB of dynamic reserve usually hurts more than gaining 1 nV/√Hz of input noise.
Spec 4 — time constant versus scan speed
A longer time constant narrows the ENBW and improves the signal-to-noise ratio, but every frequency change needs settling time. A 100-point spectrum waiting five time constants per point takes over eight minutes at 1 s, but only five seconds at 10 ms.
The OE1311 and OE2011 series cover time constants from 100 ns to 3 ks, so both ends are available: fast scans and very long averaging. If you only do slow precision measurements, short time constants are not essential — but for spectral sweeps or tracking fast changes they are decisive.
Spec 5 — channels, demodulators and interfaces
Dual channels (OE1022D, OE2052, OE2012 / OE2022) are not mainly for measuring two samples at once; they exist for differential input and two-beam cancellation, and common-mode rejection depends entirely on channel synchronism. Confirm whether the channels are truly simultaneous before ordering.
Eight demodulators (OE1311 series, OE2011 / OE2021) suit multi-frequency excitation or simultaneous harmonic measurement; the four demodulators of OE2052 paired with 400 MHz is the compromise between bandwidth and channel count. The interface decides whether automation is possible: a gigabit RJ45 is a better fit for rack-mounted unattended measurement than USB.
Three most common mistakes
- Looking only at the upper frequency limit and ignoring low-frequency drift. At the μHz level the limit is drift and 1/f, not bandwidth.
- Replacing a lock-in with a scope average or a plain preamplifier. Scope averaging improves by √N; a lock-in delivers narrow-band gain of several hundred times.
- Forgetting the chopper. Light is a DC signal, and the OE3001 optical chopper (1 Hz – 5 kHz, 1 Hz resolution, 10-slot blade) moves it to a frequency the lock-in can handle. This step is often missed.
A three-step decision path
- Step 1: fix the modulation frequency and narrow the list to two or three models with the range table above
- Step 2: fix the smallest signal and the interference level, then compare input noise and dynamic reserve
- Step 3: fix the experimental topology (single or dual channel, multi-frequency or not, automation or not), which determines channels, demodulators and interface
If two models still tie after all three steps, both are probably sufficient. At that point lead time, software maturity and service response matter more than another decimal place on a datasheet.
Models referenced in this guide
| Model | Brand | Frequency range | Input noise | Dynamic reserve | Demodulators |
|---|---|---|---|---|---|
| OE1022 Digital Lock-In Amplifier | Saiens | 1 mHz – 102 kHz | 5 nV/√Hz @997 Hz | > 120 dB | — |
| OE1022D Dual-Channel Lock-In Amplifier | Saiens | 1 mHz – 102 kHz | 5 nV/√Hz @997 Hz | — | — |
| OE1201 Digital Lock-In Amplifier | Saiens | 50 mHz – 120 kHz | Ultra-low-noise preamp | — | — |
| OE2031 / OE2041 High-Frequency Lock-In | Saiens | DC – 10 MHz / 80 MHz | 3 nV/√Hz @1 MHz | 120 dB | — |
| OE2052 Dual-Channel Lock-In Amplifier | Saiens | 1 μHz – 400 MHz | 3 nV/√Hz | — | 4 channels |
| OE3001 Optical Chopper System | Saiens | — | — | — | — |
| OE1311 / OE1351 / OE1301 Lock-In Module | Saiens | 1 μHz – 100 kHz / 500 kHz / 1 MHz | 5 nV/√Hz | > 120 dB | 8 |
| OE2011 / OE2021 Digital Lock-In Amplifier | Saiens | 10 μHz – 250 kHz / 1.5 MHz | 2.5 nV/√Hz | > 130 dB | 8 |
| OE2012 / OE2022 Dual-Channel Lock-In Amplifier | Saiens | 10 μHz – 250 kHz / 1.5 MHz | 2.5 nV/√Hz | > 130 dB | 8 |
| OE5001 Weak-Signal Teaching Kit | Saiens | — | — | — | — |
Pricing, lead time and configuration vary by project — send your requirement and an engineer replies within one business day.