Oscilloscope bandwidth, sample rate and memory: how to match three numbers
Bandwidth, sample rate and memory depth multiply together: if any one is short, the other two are wasted. Using the real specifications of nineteen oscilloscopes in our catalog, this guide shows how the three constrain each other and the four most common buying mistakes.
Three numbers that multiply
Bandwidth sets how fast an edge you can see, sample rate decides whether it is captured faithfully, and memory depth decides how long a window you can hold. They constrain each other: points per acquisition equals sample rate times time window, so when memory runs out the scope silently reduces the sample rate and the bandwidth you paid for is gone.
Bandwidth: do not pick it from the signal frequency
A square wave carries its energy in odd harmonics and only looks square after the fifth. The rule of thumb is to take at least three to five times the highest significant frequency, otherwise you are measuring the oscilloscope rather than your signal.
Rise time is the more reliable route: tr ≈ 0.35 / BW. Measuring a 500 ps edge needs more than 700 MHz; a 200 ps edge needs above 1.75 GHz. This formula is more dependable than five times the clock frequency because it maps directly onto the part of the waveform you actually care about.
- 50–200 MHz: teaching labs, power supplies and low-speed digital (SDS1000A, SDS1000E+ / F+, SDS800X EDU, SDS1000X EDU)
- 200–500 MHz: embedded work and switching-rail ripple (SDS800X HD, SDS1000X HD, SDS2000X HD / Plus)
- 1–2 GHz: high-speed digital, DDR, entry-level SerDes (SDS3000X HD, SDS5000X, SDS5000X HD, SDS6000 Pro / L)
- 8–20 GHz: PCIe and USB high-speed links, optical communications (SDS7000A, SDS8000A)
Sample rate: Nyquist is only a pass mark
Twice the bandwidth only guarantees no aliasing; it does not guarantee a usable waveform. In practice use four to five times the bandwidth. Note that headline sample rates are usually quoted for one channel or two interleaved channels, and halve when all channels are active.
Against our catalog: 1 GHz bandwidth pairs with 4 to 5 GSa/s (SDS3000X HD at 4 GSa/s, SDS5000X at 5 GSa/s); 2 GHz pairs with 10 GSa/s (SDS6000 Pro / L); 8 GHz pairs with 20 GSa/s (SDS7000A); 16 GHz pairs with 40 GSa/s in 12-bit mode on SDS8000A, or 50 GSa/s in 8-bit mode.
Memory depth: the number most often ignored
Points per acquisition equals sample rate times window length. Holding 10 GSa/s across a 100 ms window needs 1 Gpts of memory; without it the scope quietly drops the sample rate and the bandwidth you paid for never appears in long captures.
Long-memory models in the catalog: SDS5000X HD at 2.5 Gpts/ch, SDS7000A and SDS8000A at 2 Gpts/ch each, SDS3000X HD at 400 Mpts/ch and the SDS6000 series at 500 Mpts/ch. For serial-bus decode or long trend recording this figure becomes the bottleneck before bandwidth does.
Vertical resolution: when 8-bit is not enough
8-bit gives 256 levels, about 4 mV of resolution on a 1 V signal. Power-rail ripple, or a small signal riding on a large DC level, needs 10 or 12-bit. 12-bit gives 4096 levels, but only helps when the front end is quiet enough.
Noise floors can be compared directly: SDS800X HD and SDS1000X HD at 70 μVrms, SDS3000X HD at 125 μVrms, SDS5000X HD at 140 μVrms, SDS6000 Pro at 153 μVrms, SDS7000A at 300 μVrms at 8 GHz, SDS8000A at 176 μVrms at 16 GHz. High bit counts without a low noise floor lose effective bits to noise anyway.
ERES and H12 are digital filtering or averaging: they trade bandwidth for bits. A true 12-bit ADC does not. Ask whether the 12-bit is hardware or software before you buy.
Waveform capture rate: finding rare events
500,000 wfms/s means half a million waveforms per second, and the shorter the dead time, the better the odds of catching a rare glitch. In our catalog SDS7000A is rated 1,000,000 wfms/s, SDS3000X HD 890,000 and the SDS6000 series 750,000. When chasing power-up anomalies or communication errors, this figure affects whether you catch the event at all.
Four most common buying mistakes
- Comparing bandwidth but keeping a 100 MHz probe. Probe bandwidth must be at least equal to scope bandwidth, otherwise the probe limits the whole chain.
- Comparing headline sample rates without checking the rate when all channels are active.
- Treating 12-bit as a universal improvement. High bit counts need a quiet front end and enough vertical range, otherwise effective bits do not move.
- Ignoring channel count and digital inputs. A 4+16 mixed-signal model often helps embedded debug more than another 200 MHz of bandwidth.
Three steps
- Step 1: derive the bandwidth from rise time or the highest harmonic, then take one step up — bandwidth cannot be added by software later
- Step 2: multiply bandwidth by four to five for the sample rate, then multiply by the window you need for the memory depth
- Step 3: fix channel count (2, 4 or 8, plus 16 digital or not), resolution (8, 10 or 12-bit) and form factor (bench, 2U rack, handheld)
Models referenced in this guide
| Model | Brand | Bandwidth | Channels | Sample rate | Memory depth |
|---|---|---|---|---|---|
| SDS2000X HD Series | Siglent | 500 MHz | 4 + 16 | 2 GSa/s | 200 Mpts/ch |
| SDS2000X Plus Series | Siglent | 500 MHz | 4 + 16 | 2 GSa/s | 200 Mpts/ch |
| SDS3000X HD Series | Siglent | 1 GHz | 4 + 16 | 4 GSa/s | 400 Mpts/ch |
| SDS6000 Pro Series | Siglent | 2 GHz | 4 / 8 (+16 digital) | 10 GSa/s | 500 Mpts/ch |
| SDS7000A Series | Siglent | 8 GHz | 4 + 16 | 20 GSa/s | 2 Gpts/ch |
| SDS800X HD Series | Siglent | 200 MHz | 2 / 4 (+16 digital) | 2 GSa/s | 100 Mpts |
| SDS1000X-E Series | Siglent | 70 / 100 / 200 MHz | 2 / 4 (+16 digital) | 1 GSa/s | 14 Mpts |
| SHS1000X Handheld Oscilloscope | Siglent | 200 MHz | 2 + DMM | 1 GSa/s | 12 Mpts |
| SDS5000X Series Mixed Signal Oscilloscope | Siglent | 350 MHz / 500 MHz / 1 GHz | 4 + 16 digital | 5 GSa/s | 250 Mpts/CH |
| SDS5000X HD / SDS5000L High-Resolution Oscilloscope | Siglent | 350 MHz / 500 MHz / 1 GHz | 8 / 6 / 4 + EXT | 5 GSa/s | 2.5 Gpts/ch |
| SDS6000L Compact High-Resolution Oscilloscope | Siglent | 500 MHz / 1 GHz / 2 GHz | 8 / 4 + EXT | 10 GSa/s | 500 Mpts/ch |
| SDS8000A High-Bandwidth Oscilloscope | Siglent | 8 / 13 / 16 / 20 GHz | 4 + EXT | 40 GSa/s (12-bit) / 50 GSa/s (8-bit) | 2 Gpts/ch (12-bit) |
| SDS1000X HD High-Resolution Oscilloscope | Siglent | 200 MHz | 2 / 4 + 16 digital | 2 GSa/s | 100 Mpts |
| SDS1000X-U Economical 4-Channel Oscilloscope | Siglent | 100 MHz | 4 | 1 GSa/s | 14 Mpts |
| SDS1000X-C Super Phosphor Oscilloscope | Siglent | 100 / 200 MHz | 2 / 4 | 1 GSa/s | 14 Mpts |
| SDS800X EDU Teaching High-Resolution Oscilloscope | Siglent | 220 MHz | 2 / 4 + 16 digital | 2 GSa/s | 100 Mpts |
| SDS1000X EDU Teaching High-Resolution Oscilloscope | Siglent | 220 MHz | 2 / 4 + 16 digital | 2 GSa/s | 100 Mpts |
| SDS1000A Entry-Level 2-Channel Oscilloscope | Siglent | 50 / 70 / 100 / 150 / 200 MHz | 2 | 1 GSa/s | 2 Mpts |
| SDS1000E+ / SDS1000F+ Entry Oscilloscope | Siglent | 120 / 200 MHz | 2 + EXT | 1 GSa/s | 1 Mpts |
Pricing, lead time and configuration vary by project — send your requirement and an engineer replies within one business day.